Temperature cycling test is performed by Temptronic MobileTemp™ series (a ThermoStream® temperature source and a portable ThermoChamber™) with a supply of clean dry oil-free compressed air. The ThermStream® temperature source is shared with thermal vacuum test.
- MobileTemp™
- TP04300A ThermoStream®: temperature source.
- Front load style FL30 ThermoChamber™: a portable front-load style chamber.
- Compressed air supply: a compressor, an air receiver, a refrigerated air dryer, a regenerative desiccant dryer, and a dew point sensor.
ThermoStream® (updated: 10/31/2009)
Device | ThermoStream® |
Manufacturer | Temptronic Corp. |
Model | TP04300A-8C32-5 |
Function | Temperature source |
Temperature range | -80℃ to +225℃ (60 Hz) -75℃ to +225℃ (50 Hz) |
Typical temperature transition rate | -55℃ to +125℃: 10 seconds or less +125℃ to -55℃: 10 seconds or less |
DUT sensor ports | Internal diode, type T and type K thermocouple and 100 ohm platinum RTD |
Remote interface ports | IEEE-488 RS-232C serial Start test/End test/Stop on first fail (ST/ET/SFF) Ethernet |
Drivers | LabVIEW and LabWindows |
Temperature accuracy | ±1℃ |
Compressed air requirements | Filtered to 5 micro particulate contamination Oil content: < 0.01 ppm by weight filtered to 0.01 micron oil contaminant Dewpoint: <10℃ at 6.3 bars Supply pressure: 6.2 to 7.6 bars Supply flow at minimum supply pressure: 7.2L/s to 14.3 L/s (norminal 25 scfm) Humidity: 0 to 60% (45% nominal) |
Over temperature protection | +235℃ (factory set) |
Mobility | Four swivel caster wheels with locks (101.6 mm diameter) static dissipative; rear handle for ease of transport) |
Refrigerants | HCFC- and CFC-free, non-toxic, non-flammable |
Power | 200-250 VAC (230 V nominal), 50/60 Hz, 30 A, 1 phase |
Noise level | <65 dBA |
Dimension | 724 mm D x 610 mm W x 108 mm H |
Weight | 236 kg |
ThermoChamber™ (updated: 10/31/2009)
Device | Front load style ThermoChamber™ |
Manufacturer | Temptronic Corp. |
Model | FL30 |
Function | A portable thermal chamber |
Temperature range | -40℃ to +150℃ |
Temperature change range | Ambient to +125℃: 14.3 minutes +125℃ to ambient: 9.2 minutes Ambient to -40℃: 18.7 minutes -40℃ to ambient: 14 minutes |
ThermoCouple connections | 10 |
Shelves | 1 |
Linear cable feedthrough ports (port lengths 19 cm) |
2 |
Inner dimension | 510 mm D x 406 mm W x 406 mm H |
Outer dimension | 635 mm D x 510 mm W x 560 mm H |
Weight | 25.8 kg |
Compressed air supply (updated: 10/02/2010)
Device | Twin Screw Compressor |
Manufacturer | Wolf compressor Co. Ltd. |
Model | W402-10 |
Function | Compressing air |
Working pressure | 7 bar |
Max. working pressure | 15 bar |
Air flow level | 1.25 m3 / min for 7 bar, 0.87 m3 / min for 10 bar, or 0.78 m3 / min for 13 bar. |
Noise level | 68 dB |
Power | 10 HP/7.5 KW |
Voltage | 220 V/60 Hz |
Device | Air receiver |
Manufacturer | Shim Lie Hunt Co. Ltd. |
Model | Stand |
Function | Storing compressed air |
Max. A.W.P. | 10 kg cm-2 |
Hyadrostic test | 16 kg cm-2 |
Volume | 0.304 m3 (304 L) |
Designed temperature | 40℃ |
Valve seat diameter | 32 mm |
Material | SS400 |
Device | Refrigerative Air Dryer |
Manufacturer | Yu-Chiao Machinery Co. Ltd.(祐僑機械股份有限公司) |
Model | YC10AS |
Function | Filter out water from compressed air |
Used pressure | 10 kg cm-2 |
Input temperature | 80℃ |
HP | 1/3 |
Outlet diameter | 3/4" |
scfm | 43 |
Dew point | 2 to 8 oC (1.7 oC by design) |
Dispose wind | 1.2 nm3 min-1 |
Condenser | R134 |
Voltage | 220 V/60 Hz |
Dimension (mm) | 700 (L) x 400 (W) x 650 (H) |
Weight | 57 kg |
Device | Regenerative Desiccant Air Dryer |
Manufacturer | Hyper Dryer Machinery Co. Ltd.(高乾機械股份有限公司) |
Model | F-10M |
Function | Filter out water from compressed air |
Used pressure | 0.7 to 1 MPa (10 Kg/cm2) |
Input temperature | <45℃ |
HP | 1 |
Outlet diameter | 3/4" |
Max flow rate | 1.3 m3/min |
Dew point | -40 to -70 oC |
Voltage | 220 V/60 Hz and 1A |
Dimension (mm) | 510 (L) x 500 (W) x 1475 (H) |
Device | Dew point sensor mobile |
Manufacturer | BEKO Technologies GmbH |
Model | METPOINT DP 207 |
Function | Measure dew point of compressed air. |
Dew point pressure | -80℃ to +45℃ td |
Humidity | 0 to 100 %RH |
Operating temperature | -20℃ to +70℃ |
Pressure range | Installation without measuring chamber: –1 to 50 bar Installation with a measuring chamber: up to 16 bar |
Power supply | Internal Internal NiMH battery for approx. 15 h of operation, indication ”Low Bat. Max 2 h”, residual operating time, charging time: max. 2 h |
Power supply unit | Input: 110 - 240 VAC, 50 - 60 Hz, 0.2 A Output: 24 VDC, 300 mA, safety category 2, for application in dry places only |
Weight | 340 g |